RED01
Relia D
Device Survival Probability
ADDR01_ADDR03_ADDR06_ADD01_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | at 162.0 mo | |||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | ||||||||
% | 99.9 | 99.9 | 99.8 | 99.6 | 99.4 | 98.9 | 97.9 | 94.9 | 87.2 | 73.5 | 59.4 | 43.8 | 29.0 | 17.1 | ||||||||
# | 393196 | 365311 | 338692 | 312687 | 288595 | 263716 | 235296 | 200861 | 157429 | 106364 | 63561 | 29901 | 8351 | 732 |
- Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
- Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Product Characteristics
Generator Type | IPG |
---|---|
NBD Code | DDD |
Max. Delivered Energy | N/A |
Connector Style | IS-1 UNI/BI |
---|---|
Serial Number Prefix | NWW |
Mass | 27.1 g |
Volume | 12.1 cc |
Waveform | N/A |
X-Ray ID | PWB |
IS-1 UNI/BI
Estimated Longevity
Amplitude Setting | 500 Lead Ω | 1000 Lead Ω |
---|---|---|
Low 2.5 V (A, RV) | 9.1 | 10.3 |
Nominal 3.5 V (A, RV) | 7.2 | 9.0 |
High 5.0 V (A, RV) | 5.1 | 7.0 |
Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; EGM Prestorage ON ; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; EGM Prestorage ON Time 1 ; Atrial Fib Percent of Time 0 %Hide this content
Distribution Data
US Market Release | |
---|---|
CE Approval Date | 2008-05-07 |
Registered USA Implants | 1 |
Estimated Active USA Implants | 0 |
Normal Battery Depletions |
US Market Release | |
---|---|
CE Approval Date | 2008-05-07 |
Estimated WW Distribution | 63910 |
Normal Battery Depletions | 50 |
EOS Indication
The prolonged service period is 3 months after RRT/ERI and assumes 100% pacing at VVI/65 and the listed parameter values. Reprogramming of parameters may shorten the duration of the prolonged service period.
Data as of December 1, 2023