6935
Sprint Quattro Secure S
ALL

Device Survival Probability
6935_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
99.5 | 99.2 | 98.9 | 98.6 | 98.4 | 97.9 | 97.3 | 96.4 | 95.4 | 94.6 | 94.6 | ||||||||||
# | 2314 | 1908 | 1566 | 1259 | 1060 | 912 | 751 | 549 | 321 | 191 | 77 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 2817 |
---|---|
Cumulative Months of Follow-Up | 145962 |
Number of Leads Active in Study | 810 |

US Acute Lead Observations
Cardiac Perforation | 26 |
---|---|
Conductor Fracture | 4 |
Extra Cardiac Stimulation | 1 |
Failure to Capture | 28 |
Failure to Sense | 14 |
Impedance Out of Range | 26 |
Insulation Breach | 1 |
Lead Dislodgement | 65 |
Oversensing | 63 |
Unspecified Clinical Failure | 5 |

Product Characteristics
Lead Type | Defibrillation |
---|---|
Type and/or Fixation | Active Screw In |
Lead Function | Pacing/Sensing and Defibrillation |
Lead Placement | Transvenous |
Lead Tip Location | Right Ventricle |
Pace/Sense Polarity | True Bipolar/One Coil |
Serial Number Prefix | TAU |
---|---|
Lead Body Diameter | 8,2 FR |
Insulation | Silicone with Polyurethane Overlay |
Pace/Sense Connector Style | IS-1 BI |
High Voltage Connector | DF-1 |
Steroid | Yes |
Data as of December 16, 2021