P1501DR EnRhythm DR
Implantable Pulse Generator (IPG)
P1501DR EnRhythm DR
Information

Device Survival Probability

P1501DR_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr at 174.0 mo            
% Excluding Normal Battery Depletion 100.0 100.0 99.7 98.1 93.6 86.6 80.8 76.7 74.2 73.6 73.4 73.3 73.0 72.9 72.8            
% Including Normal Battery Depletion 99.9 99.8 99.3 97.1 91.1 80.3 66.5 51.2 36.6 23.8 16.2 13.6 12.7 12.0 10.8            
#   94969 88744 82391 74749 64545 51290 37804 25073 15196 8340 4733 3146 2037 796 155            
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

Product Characteristics

Generator Type IPG
NBD Code DDDRP
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix PNP
Mass 21.0 g
Volume 13.0 cc
Waveform N/A
X-Ray ID PNP
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 9.7 11.2
Nominal 3.5 V (A, RV) 7.3 9.4
High 5.0 V (A, RV) 4.9 7.1
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %; These estimates are adjusted to reflect changes addressed by February 2010 Advisory "Low Battery Voltage Displayed at Device Interrogation".Hide this content

Information

Distribution Data

US Market Release2005-05-05
CE Approval Date2004-08-13
Registered USA Implants109982
Estimated Active USA Implants7702
Normal Battery Depletions17538
US Market Release2005-05-05
CE Approval Date2004-08-13
Estimated WW Distribution141130
Normal Battery Depletions18121
Information

Malfunctions (USA)

Therapy Function Not Compromised 15113
Battery 14982
Electrical Component 59
Electrical Interconnect 2
Possible Early Battery Depletion 69
Other 1
Therapy Function Compromised 55
Battery 6
Electrical Component 38
Electrical Interconnect 4
Possible Early Battery Depletion 2
Other 5
Information

Malfunctions (WW)

Therapy Function Not Compromised 17883
Battery 17670
Electrical Component 116
Electrical Interconnect 5
Possible Early Battery Depletion 88
Software/Firmware 1
Other 3
Therapy Function Compromised 77
Battery 10
Electrical Component 52
Electrical Interconnect 5
Possible Early Battery Depletion 5
Other 5
Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Elective Replacement (ERI)

Telemetry indication. Rate and Mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of December 1, 2023

www.medtronic.com/productperformance 951051