P1501DR
EnRhythm DR

Device Survival Probability
P1501DR_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | at 129.0 mo | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
99.9 | 99.9 | 99.7 | 98.0 | 93.6 | 86.9 | 81.5 | 78.0 | 76.0 | 75.6 | 75.6 | ||||||||||
% | ![]() |
99.6 | 99.5 | 99.0 | 96.5 | 90.1 | 77.9 | 62.4 | 45.3 | 29.6 | 15.6 | 2.7 | ||||||||||
# | 94470 | 88156 | 82115 | 75107 | 65154 | 51147 | 36780 | 23442 | 13337 | 4914 | 272 |
- Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
- Excluding Normal Battery Depletion – This is the malfunction free survival curve.

Product Characteristics
Generator Type | IPG |
---|---|
NBD Code | DDDRP |
Max. Delivered Energy | N/A |
Connector Style | IS-1 UNI/BI |
---|---|
Serial Number Prefix | PNP |
Mass | 21.0 g |
Volume | 13.0 cc |
Waveform | N/A |
X-Ray ID | PNP |
IS-1 UNI/BI

Estimated Longevity
Amplitude Setting | 500 Lead Ω | 1000 Lead Ω |
---|---|---|
Low 2.5 V (A, RV) | 9.7 | 11.2 |
Nominal 3.5 V (A, RV) | 7.3 | 9.4 |
High 5.0 V (A, RV) | 4.9 | 7.1 |
Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %; These estimates are adjusted to reflect changes addressed by February 2010 Advisory "Low Battery Voltage Displayed at Device Interrogation".Hide this content

Distribution Data
US Market Release | 2005-05-05 |
---|---|
CE Approval Date | 2004-08-13 |
Registered USA Implants | 109829 |
Estimated Active USA Implants | 17101 |
Normal Battery Depletions | 17172 |
US Market Release | 2005-05-05 |
---|---|
CE Approval Date | 2004-08-13 |
Estimated WW Distribution | 142034 |
Normal Battery Depletions | 17774 |

Malfunctions (USA)
Therapy Function Not Compromised | 15019 |
---|---|
Battery | 14888 |
Electrical Component | 59 |
Electrical Interconnect | 2 |
Possible Early Battery Depletion | 69 |
Other | 1 |
Therapy Function Compromised | 55 |
Battery | 6 |
Electrical Component | 38 |
Electrical Interconnect | 4 |
Possible Early Battery Depletion | 2 |
Other | 5 |

Malfunctions (WW)
Therapy Function Not Compromised | 17786 |
---|---|
Battery | 17576 |
Electrical Component | 115 |
Electrical Interconnect | 4 |
Possible Early Battery Depletion | 87 |
Software/Firmware | 1 |
Other | 3 |
Therapy Function Compromised | 76 |
Battery | 10 |
Electrical Component | 51 |
Electrical Interconnect | 5 |
Possible Early Battery Depletion | 5 |
Other | 5 |
Data as of October 29, 2020