6947
Sprint Quattro Secure
ALL
Device Survival Probability
6947_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | 14 yr | 15 yr | 16 yr | 17 yr | 18 yr | 19 yr | 20 yr | |||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | 99.5 | 99.3 | 99.0 | 98.7 | 98.2 | 97.9 | 97.5 | 97.1 | 96.7 | 96.1 | 95.9 | 95.5 | 95.0 | 94.6 | 94.3 | 92.5 | 92.5 | 91.0 | 88.9 | 86.2 | ||
# | 3307 | 2907 | 2552 | 2262 | 2027 | 1788 | 1545 | 1385 | 1239 | 1085 | 911 | 720 | 561 | 399 | 241 | 159 | 122 | 109 | 82 | 50 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 4600 |
---|---|
Cumulative Months of Follow-Up | 302679 |
Number of Leads Active in Study | 529 |
US Acute Lead Observations
Cardiac Perforation | 29 |
---|---|
Conductor Fracture | 26 |
Extra Cardiac Stimulation | 2 |
Failure to Capture | 83 |
Failure to Sense | 36 |
Impedance Out of Range | 61 |
Insulation Breach | 4 |
Lead Dislodgement | 125 |
Oversensing | 141 |
Unspecified Clinical Failure | 20 |
Product Characteristics
Lead Type | Defibrillation |
---|---|
Type and/or Fixation | Active Screw In |
Lead Function | Pacing/Sensing and Defibrillation |
Lead Placement | Transvenous |
Lead Tip Location | Right Ventricle |
Pace/Sense Polarity | True Bipolar/Two Coils |
Serial Number Prefix | TDG |
---|---|
Lead Body Diameter | 8.6 FR |
Insulation | Silicone with Polyurethane Overlay |
Pace/Sense Connector Style | IS-1 BI |
High Voltage Connector | DF-1 |
Steroid | Yes |
Data as of November 20, 2024