C4TR01
Consulta CRT-P

Device Survival Probability
C2TR01_C3TR01_C4TR01_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | at 93.0 mo | ||||
---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
100.0 | 100.0 | 100.0 | 100.0 | 99.9 | 99.9 | 99.9 | 99.9 | ||
% | ![]() |
99.9 | 99.8 | 99.5 | 98.1 | 95.8 | 91.8 | 84.2 | 58.7 | ||
# | 26671 | 23951 | 21390 | 18233 | 14209 | 9522 | 3922 | 374 |
- Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
- Excluding Normal Battery Depletion – This is the malfunction free survival curve.

Product Characteristics
Generator Type | CRT-Pacing |
---|---|
NBD Code | OAE-DDDR |
Max. Delivered Energy | N/A |
Connector Style | CRT-P Style |
---|---|
Serial Number Prefix | PVX |
Mass | 26 g |
Volume | 15 cc |
Waveform | N/A |
X-Ray ID | PVX |
CRT-P Style

Estimated Longevity
Amplitude Setting | 500 Lead Ω | 1000 Lead Ω |
---|---|---|
Low 2.5 V (A, RV) | 8.7 | 10.7 |
Nominal 3.5 V (A, RV) | 6.0 | 8.2 |
High 5.0 V (A, RV) | 3.3 | 5.1 |
Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Left Ventricle Pulse Width 0.4 ms; Left Ventricle Pulse Amplitude same as A, RV; Left Ventricle Percent Paced 100 %; Left Ventricle Lead Impedance same as A, RV; EGM Prestorage ON ; EGM Prestorage ON Time 1 ; Atrial Fib Percent of Time 0 %Hide this content

Distribution Data
US Market Release | 2011-03-22 |
---|---|
CE Approval Date | |
Registered USA Implants | 23541 |
Estimated Active USA Implants | 14824 |
Normal Battery Depletions | 966 |
US Market Release | 2011-03-22 |
---|---|
CE Approval Date | |
Estimated WW Distribution | 23817 |
Normal Battery Depletions | 966 |
Data as of October 29, 2020