4074 CapSure Sense
Pacing
4074 CapSure Sense
VENTRICULAR PLACEMENT
Information

Device Survival Probability

4074_VEN_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr                  
% Survival 99.5 99.4 99.3 99.3 99.3 99.3 99.3 99.3 99.3 98.7 98.7 98.7                  
#   982 815 676 548 376 274 210 182 159 132 83 52                  

Product Surveillance Registry Results

Number of Leads Enrolled in Study 1124
Cumulative Months of Follow-Up 60823
Number of Leads Active in Study 378
Information 8

Qualifying Complications

Conductor Fracture 1
Failure to Capture 2
Impedance Out of Range 1
Insulation (not further defined) 1
Lead Dislodgement 2
Other 1
ATRIAL PLACEMENT
Information

Device Survival Probability

4074_ATR_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr                  
% Survival 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1                  
#   214 205 198 183 167 158 148 134 122 112 74 60                  

Product Surveillance Registry Results

Number of Leads Enrolled in Study 227
Cumulative Months of Follow-Up 23130
Number of Leads Active in Study 104
Information 2

Qualifying Complications

Failure to Sense 1
Lead Dislodgement 1
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Cardiac Perforation 21
Conductor Fracture 1
Extra Cardiac Stimulation 3
Failure to Capture 73
Failure to Sense 3
Impedance Out of Range 4
Lead Dislodgement 91
Oversensing 4
Information

Distribution Data

US Market Release2002-06-23
CE Approval Date2002-02-01
Registered USA Implants119708
Estimated Active USA Implants70043
US Market Release2002-06-23
CE Approval Date2002-02-01
Estimated WW Distribution792742
Information

USA Returned Product Analysis

Conductor Fracture 9
Insulation Breach 35
Crimp/Weld/Bond 0
Other 0
Information

WW Returned Product Analysis

Conductor Fracture 15
Insulation Breach 38
Crimp/Weld/Bond 0
Other 10
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Tines
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity Bipolar
More
Serial Number Prefix BBD
Lead Body Diameter 5.3 FR
Insulation Polyurethane (outer) Silicone (inner)
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of January 31, 2018

www.medtronic.com/productperformance 601051