6935M
Sprint Quattro Secure S
ALL

Device Survival Probability
6935M_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | ||||
---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
99.6 | 99.5 | 99.2 | 98.9 | 98.4 | 97.7 | 97.5 | 97.5 | ||
# | 5712 | 4615 | 3625 | 2768 | 1746 | 867 | 353 | 52 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 7355 |
---|---|
Cumulative Months of Follow-Up | 274486 |
Number of Leads Active in Study | 4056 |

US Acute Lead Observations
Cardiac Perforation | 141 |
---|---|
Conductor Fracture | 16 |
Extra Cardiac Stimulation | 26 |
Failure to Capture | 291 |
Failure to Sense | 91 |
Impedance Out of Range | 85 |
Insulation Breach | 2 |
Lead Dislodgement | 506 |
Oversensing | 259 |

Product Characteristics
Lead Type | Defibrillation |
---|---|
Type and/or Fixation | Active Screw In |
Lead Function | Pacing/Sensing and Defibrillation |
Lead Placement | Transvenous |
Lead Tip Location | Right Ventricle |
Pace/Sense Polarity | True Bipolar/One Coil |
Serial Number Prefix | TDL |
---|---|
Lead Body Diameter | 2.8 mm (8.4 Fr) |
Insulation | Silicone with Polyurethane Overlay |
Pace/Sense Connector Style | DF4-LLHO |
High Voltage Connector | DF4-LLHO |
Steroid | Yes |
Data as of December 16, 2021