ADVDD01 Adapta VDD
Implantable Pulse Generator (IPG)
ADVDD01 Adapta VDD
Information

Device Survival Probability

ADVDD01_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr at 109.0 mo
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0
% Including Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 97.3 87.0 64.7 50.7 50.7
#   1223 1144 1017 921 826 703 513 248 111 100
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Customer Communications For This Model

Information

Product Characteristics

Generator Type IPG
NBD Code VDD
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix NWG, PWG
Mass 23.6 g
Volume 11.1 cc
Waveform N/A
X-Ray ID PWB, NWG
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 6.2 | 7.9* 6.5 | 8.5*
Nominal 3.5 V (A, RV) 5.5 | 7.0* 6.2 | 8.1*
High 5.0 V (A, RV) 4.4 | 5.4* 5.4 | 6.7*
Footnote

*The second longevity numbers provided apply to the N-series Adapta ADVDD01 product. The N-series products have increased longevity because of improved electronic circuitry. Click on "Other Articles" for more information on the N-series longevity. Longevity estimates based on the following device usage. Pace/Sense Mode VDD ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; EGM Prestorage ON ; Upper Sensing Rate 130 ppm; EGM Prestorage ON Time 1 ; Atrial Fib Percent of Time 0 %; Hide this content

Information

Distribution Data

US Market Release2006-07-17
CE Approval Date2005-09-20
Registered USA Implants1421
Estimated Active USA Implants648
Normal Battery Depletions94
US Market Release2006-07-17
CE Approval Date2005-09-20
Estimated WW Distribution73091
Normal Battery Depletions398
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised 1
Electrical Interconnect 1
Therapy Function Compromised 2
Electrical Component 1
Electrical Interconnect 1
Information

EOS Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Recommended Replacement (RRT)

Telemetry indication. Rate and mode change to 65 ppm and VVI respectively (VOO/65 with magnet).

Data as of October 29, 2020

www.medtronic.com/productperformance 776050