SPSR01 Sphera SR MRI
Implantable Pulse Generator (IPG)
SPSR01 Sphera SR MRI
Information

Device Survival Probability

SPHERA_SR_SURV
Loading...










    at 0.0 mo                  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

Product Characteristics

Generator Type IPG
NBD Code AAIR/VVIR, AAI/VVI
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix FNG
Mass 21.5 g
Volume 9.7 cc
Waveform N/A
X-Ray ID PHX
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 9.4 10.3
Nominal 3.5 V (A, RV) 8.1 9.7
High 5.0 V (A, RV) 6.1 7.8
Footnote

Longevity estimates based on the following device usage.   Pace/Sense Mode VVI ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Hide this content

Information

Distribution Data

US Market Release2017-08-03
CE Approval Date2017-06-16
Registered USA Implants
Estimated Active USA Implants
Normal Battery Depletions
US Market Release2017-08-03
CE Approval Date2017-06-16
Estimated WW Distribution34100
Normal Battery Depletions
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised
Therapy Function Compromised
Information

EOL Indicator

180 days from RRT

Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.5 V, or battery voltage less than or equal to 2.59 V and battery impedance at 3000 ohms

Data as of December 16, 2021

www.medtronic.com/productperformance 851050