DTBB1D4 Viva S
Cardiac Resynchronization Therapy - Defibrillation (CRT-Defib)
DTBB1D4 Viva S
Information

Device Survival Probability

BRAVA_VIVA_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr at 62.0 mo        
% Excluding Normal Battery Depletion 100.0 100.0 99.9 99.9 99.9 99.9        
% Including Normal Battery Depletion 99.9 99.8 99.0 97.1 93.3 92.1        
#   89299 74382 52084 29075 4221 1178        
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.

Performance Notes For This Model

    • There are no performance notes for this model.
Information

Product Characteristics

Generator Type CRT-Defib
NBD Code DDE-DDDR
Max. Delivered Energy 36 J
More
Connector Style DF4 CRT-D Style
Serial Number Prefix BLN
Mass 80 g
Volume 35 cc
Waveform Biphasic
X-Ray ID PXT
Hide this content
DF4 CRT-D Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 3.6 4.3 4.9 5.3
Quarterly 4.9 6.3 7.8 8.8
Semi-annual 5.2 6.6 8.5 9.6
Footnote

Longevity estimates based on the following device usage. Pacing Mode DDD; Atrial Lower Pacing Rate 60ppm; Ventricular Lower Pacing Rate 60ppm; Atrial Pulse Width 0.4ms; Ventricular Pulse Width 0.4ms; Atrial Pulse Amplitude 3V; Ventricular Pulse Amplitude 3V; Atrial Impedance 510ohms; Ventricle Impedance 510ohms; Atrial Sensing Rate 70bpm; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Integrated bi-ventricular pacing RV+LV; Left Ventricle Impedance 510ohms; Left Ventricular Pulse Width 0.4ms; Left Ventricular Pulse Amplitude 3V; Optivol ON; Atrial Fib percentage of time 0%; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

Distribution Data

US Market Release2013-01-29
CE Approval Date
Registered USA Implants4448
Estimated Active USA Implants3798
Normal Battery Depletions84
US Market Release2013-01-29
CE Approval Date
Estimated WW Distribution5003
Normal Battery Depletions84
Information

Malfunctions (USA)

Therapy Function Not Compromised 3
Battery 1
Electrical Component 1
Other 1
Therapy Function Compromised 2
Battery 2
Information

Malfunctions (WW)

Therapy Function Not Compromised 4
Battery 1
Electrical Component 2
Other 1
Therapy Function Compromised 2
Battery 2
Information

EOS Indication

From the point that the RRT is set, the pacemaker will operate for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.73V on 3 consecutive daily automatic measurments.

Information

Charge Time

BLACKWELL_CRT-D_CHRG
Loading...

Data as of October 12, 2018

www.medtronic.com/productperformance 626050