5054
CapSure Z Novus
VENTRICULAR PLACEMENT

Device Survival Probability
5054_VEN_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | |||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
99.3 | 99.1 | 98.8 | 98.4 | 98.4 | 97.0 | 97.0 | 96.4 | 96.4 | 96.4 | 96.4 | 96.4 | |||||||||
# | 474 | 391 | 304 | 263 | 229 | 191 | 167 | 142 | 110 | 91 | 69 | 51 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 989 |
---|---|
Cumulative Months of Follow-Up | 34889 |
Number of Leads Active in Study | 27 |
ATRIAL PLACEMENT

Device Survival Probability
5054_ATR_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | 14 yr | 15 yr | at 186.0 mo | |||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
99.5 | 99.5 | 99.5 | 99.5 | 99.5 | 99.5 | 99.5 | 99.5 | 99.5 | 99.5 | 99.5 | 99.5 | 99.5 | 99.5 | 99.5 | 99.5 | |||||
# | 411 | 391 | 358 | 322 | 289 | 252 | 219 | 186 | 153 | 129 | 108 | 93 | 75 | 65 | 56 | 53 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 426 |
---|---|
Cumulative Months of Follow-Up | 41161 |
Number of Leads Active in Study | 41 |

US Acute Lead Observations
Cardiac Perforation | 2 |
---|---|
Conductor Fracture | 2 |
Failure to Capture | 23 |
Impedance Out of Range | 4 |
Insulation Breach | 1 |
Lead Dislodgement | 30 |
Unspecified Clinical Failure | 9 |

Product Characteristics
Lead Type | Pacing |
---|---|
Type and/or Fixation | Tines |
Lead Function | Pacing/Sensing |
Lead Placement | Transvenous |
Lead Tip Location | Right Ventricle |
Pace/Sense Polarity | Bipolar |
Serial Number Prefix | LEH |
---|---|
Lead Body Diameter | 6.0 FR |
Insulation | Silicone |
Pace/Sense Connector Style | IS-1 BI |
High Voltage Connector | n/a |
Steroid | Yes |
Data as of December 16, 2021