KSR903 Kappa 900 SR
Implantable Pulse Generator (IPG)
KSR903 Kappa 900 SR
Information

Device Survival Probability

KSR901_KSR903_KSR906_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr at 119.0 mo                      
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 99.9 99.9 99.9 99.9 99.9                      
% Including Normal Battery Depletion 99.4 98.9 98.5 97.2 95.5 91.9 80.0 49.2 16.7 3.0                      
#   28635 24217 20635 17358 14465 11822 8728 4392 1109 111                      
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Advisories For This Model

  • There are no advisories for this model.

Performance Notes For This Model

    • There are no performance notes for this model.
Information

Product Characteristics

Generator Type IPG
NBD Code SSIR
Max. Delivered Energy N/A
More
Connector Style 3.2mm Long Pin
Serial Number Prefix PLG
Mass 22.5 g
Volume 10.5 cc
Waveform N/A
X-Ray ID PLG
Hide this content
3.2mm Long Pin
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 7.3 7.9
Nominal 3.5 V (A, RV) 6.4 7.4
High 5.0 V (A, RV) 4.9 6.1
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode VVI ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppmHide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release2002-01-09
CE Approval Date2001-09-28
Registered USA Implants1381
Estimated Active USA Implants87
Normal Battery Depletions166
US Market Release2002-01-09
CE Approval Date2001-09-28
Estimated WW Distribution13713
Normal Battery Depletions1153
Information

Malfunctions (USA)

Therapy Function Not Compromised 1
Electrical Component 1
Therapy Function Compromised 0
Information

Malfunctions (WW)

Therapy Function Not Compromised 2
Electrical Component 1
Possible Early Battery Depletion 1
Therapy Function Compromised 3
Electrical Component 2
Electrical Interconnect 1
Information

Elective Replacement (ERI)

Telemetry indication. Rate and Mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of February 3, 2017