KSR701 Kappa 700 SR
Implantable Pulse Generator (IPG)
KSR701 Kappa 700 SR
Information

Device Survival Probability

KSR700_KSR701_KSR703_KSR706_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr at 107.0 mo                        
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 99.9 99.9 99.8                        
% Including Normal Battery Depletion 99.5 99.1 98.3 96.7 93.6 88.1 74.0 43.3 1.7                        
#   42989 36388 30508 25423 20691 16234 11242 5052 259                        
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Advisories For This Model

  • There are no advisories for this model.

Performance Notes For This Model

Information

Product Characteristics

Generator Type IPG
NBD Code SSIR
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix PHT
Mass 21.5 g
Volume 9.7 cc
Waveform N/A
X-Ray ID PHT
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 7.4 7.9
Nominal 3.5 V (A, RV) 6.5 7.5
High 5.0 V (A, RV) 4.9 6.2
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode VVI ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppmHide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release1999-01-29
CE Approval Date1998-03-20
Registered USA Implants48451
Estimated Active USA Implants2645
Normal Battery Depletions5185
US Market Release1999-01-29
CE Approval Date1998-03-20
Estimated WW Distribution90107
Normal Battery Depletions5802
Information

Malfunctions (USA)

Therapy Function Not Compromised 3
Electrical Component 1
Electrical Interconnect 1
Possible Early Battery Depletion 1
Therapy Function Compromised 19
Electrical Component 2
Electrical Interconnect 17
Information

Malfunctions (WW)

Therapy Function Not Compromised 17
Electrical Component 10
Electrical Interconnect 4
Possible Early Battery Depletion 2
Software/Firmware 1
Therapy Function Compromised 73
Electrical Component 4
Electrical Interconnect 69
Information

Elective Replacement (ERI)

Telemetry indication. Rate and mode change to 65 ppm and VVI respectively (VOO/65 with magnet).

Data as of August 3, 2016