KDR700 Kappa 700 DR
Implantable Pulse Generator (IPG)
KDR700 Kappa 700 DR
Information

Device Survival Probability

KD700_KDR700_KDR701_KDR703_KDR706_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr at 116.0 mo
% Excluding Normal Battery Depletion 100.0 100.0 99.9 99.9 99.9 99.9 99.7 99.6 99.4 99.4
% Including Normal Battery Depletion 99.5 99.4 99.0 98.2 96.6 93.0 82.0 55.8 23.8 3.4
#   180086 165059 150466 135831 121509 105917 83642 48032 14051 1419
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Advisories For This Model

  • There are no advisories for this model.
Information

Product Characteristics

Generator Type IPG
NBD Code DDD/RO
Max. Delivered Energy N/A
More
Serial Number Prefix PGT
Mass 27.0 g
Volume 11.9 cc
Waveform N/A
X-Ray ID PGT
Hide this content
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 7.7 8.6
Nominal 3.5 V (A, RV) 6.3 7.7
High 5.0 V (A, RV) 4.4 6.0
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release
CE Approval Date
Registered USA Implants15
Estimated Active USA Implants1
Normal Battery Depletions4
US Market Release
CE Approval Date
Estimated WW Distribution2773
Normal Battery Depletions29
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised 0
Therapy Function Compromised 1
Electrical Interconnect 1
Information

Elective Replacement (ERI)

Replacement message on programmer. Nonmagnet mode VVI 65 ppm Magnet mode VOO 65 ppm

Data as of August 3, 2014

www.medtronic.com/productperformance 326050