Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Longevity estimates based on the following device usage.
Pacing Mode DDD;
Atrial Lower Pacing Rate 60ppm;
Ventricular Lower Pacing Rate 60ppm;
Atrial Pulse Width 0.4ms;
Ventricular Pulse Width 0.4ms;
Atrial Pulse Amplitude 3V;
Ventricular Pulse Amplitude 3V;
Atrial Impedance 510ohms;
Ventricle Impedance 510ohms;
Atrial Sensing Rate 70bpm;
Ventricle Sensing Rate 70bpm;
EGM prestorage OFF;
Integrated bi-ventricular pacing RV+LV;
Left Ventricle Impedance 510ohms;
Left Ventricular Pulse Width 0.4ms;
Left Ventricular Pulse Amplitude 3V;
Optivol ON;
Atrial Fib percentage of time 0%;
Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation.
Hide this content
Distribution Data
US Market Release
2008-09-15
CE Approval Date
Registered USA Implants
65130
Estimated Active USA Implants
5026
Normal Battery Depletions
18955
US Market Release
2008-09-15
CE Approval Date
Estimated WW Distribution
66577
Normal Battery Depletions
18982
Malfunctions (USA)
Therapy Function Not Compromised
573
Battery
2
Electrical Component
67
Electrical Interconnect
1
Possible Early Battery Depletion
496
Software/Firmware
6
Other
1
Therapy Function Compromised
31
Battery
5
Electrical Component
26
Malfunctions (WW)
Therapy Function Not Compromised
576
Battery
2
Electrical Component
69
Electrical Interconnect
1
Possible Early Battery Depletion
497
Software/Firmware
6
Other
1
Therapy Function Compromised
31
Battery
5
Electrical Component
26
EOS Indication
From the point that the RRT is set, the pacemaker will operate for approximately three months for typical pacemaker configurations during the normal operating life.
Recommended Replacement (RRT)
Battery voltage less than 2.63 V or Charge Time greater than 16 seconds.