Device Survival Probability
CONSULTA_CRT_SURV
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | 100.0 | 99.7 | 99.3 | 98.8 | 98.7 | 98.5 | 98.4 | 98.4 | 98.4 | 98.4 | 98.3 | 98.3 | 98.3 | |||||||||
% | 99.7 | 98.4 | 93.3 | 81.3 | 58.9 | 29.8 | 20.2 | 18.2 | 17.5 | 16.9 | 16.5 | 16.3 | 15.2 | |||||||||
# | 56118 | 50205 | 42857 | 33041 | 19346 | 7437 | 4019 | 3271 | 2912 | 2636 | 2080 | 1400 | 122 |
- Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
- Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Product Characteristics
Generator Type | CRT-Defib |
---|---|
NBD Code | DDE-DDDR |
Max. Delivered Energy | 35 J |
Connector Style | DF4 CRT-D Style |
---|---|
Serial Number Prefix | PUC |
Mass | 73 g |
Volume | 40 cc |
Waveform | Biphasic |
X-Ray ID | PUG |
Estimated Longevity
Charging Frequency | 100% Pacing | 50% Pacing | 15% Pacing | 100% Sensing |
---|---|---|---|---|
Monthly | 3.2 | 3.8 | 4.4 | 4.7 |
Quarterly | 4.4 | 5.5 | 6.8 | 7.5 |
Semi-annual | 4.8 | 6.2 | 7.9 | 9.0 |
Longevity estimates based on the following device usage. Pacing Mode DDD; Atrial Lower Pacing Rate 60ppm; Ventricular Lower Pacing Rate 60ppm; Atrial Pulse Width 0.4ms; Ventricular Pulse Width 0.4ms; Atrial Pulse Amplitude 3V; Ventricular Pulse Amplitude 3V; Atrial Impedance 510ohms; Ventricle Impedance 510ohms; Atrial Sensing Rate 70bpm; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Integrated bi-ventricular pacing RV+LV; Left Ventricle Impedance 510ohms; Left Ventricular Pulse Width 0.4ms; Left Ventricular Pulse Amplitude 3V; Optivol ON; Atrial Fib percentage of time 0%; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content
Distribution Data
US Market Release | 2012-01-09 |
---|---|
CE Approval Date | |
Registered USA Implants | 2048 |
Estimated Active USA Implants | 259 |
Normal Battery Depletions | 722 |
US Market Release | 2012-01-09 |
---|---|
CE Approval Date | |
Estimated WW Distribution | 2190 |
Normal Battery Depletions | 724 |
EOS Indication
From the point that the RRT is set, the pacemaker will operate for approximately three months for typical pacemaker configurations during the normal operating life.