6947
Sprint Quattro Secure
ALL
Device Survival Probability
6947_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | 14 yr | 15 yr | 16 yr | 17 yr | 18 yr | 19 yr | ||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | 99.5 | 99.3 | 99.0 | 98.7 | 98.2 | 97.9 | 97.5 | 97.1 | 96.7 | 96.1 | 95.8 | 95.4 | 94.9 | 94.3 | 93.9 | 91.5 | 91.5 | 90.7 | 88.2 | |||
# | 3297 | 2899 | 2546 | 2255 | 2022 | 1782 | 1533 | 1372 | 1223 | 1049 | 861 | 638 | 456 | 280 | 177 | 139 | 117 | 98 | 59 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 4568 |
---|---|
Cumulative Months of Follow-Up | 294892 |
Number of Leads Active in Study | 618 |
US Acute Lead Observations
Cardiac Perforation | 29 |
---|---|
Conductor Fracture | 26 |
Extra Cardiac Stimulation | 2 |
Failure to Capture | 83 |
Failure to Sense | 36 |
Impedance Out of Range | 61 |
Insulation Breach | 4 |
Lead Dislodgement | 124 |
Oversensing | 141 |
Unspecified Clinical Failure | 20 |
Product Characteristics
Lead Type | Defibrillation |
---|---|
Type and/or Fixation | Active Screw In |
Lead Function | Pacing/Sensing and Defibrillation |
Lead Placement | Transvenous |
Lead Tip Location | Right Ventricle |
Pace/Sense Polarity | True Bipolar/Two Coils |
Serial Number Prefix | TDG |
---|---|
Lead Body Diameter | 8.6 FR |
Insulation | Silicone with Polyurethane Overlay |
Pace/Sense Connector Style | IS-1 BI |
High Voltage Connector | DF-1 |
Steroid | Yes |
Data as of December 1, 2023