6944
Sprint Quattro
ALL
Device Survival Probability
6944_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | 100.0 | 99.8 | 99.2 | 97.3 | 94.8 | 91.7 | 91.1 | 90.6 | 89.9 | 89.9 | 89.0 | 88.1 | 85.7 | |||||||||
# | 502 | 418 | 352 | 290 | 228 | 191 | 165 | 145 | 132 | 115 | 101 | 79 | 53 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 638 |
---|---|
Cumulative Months of Follow-Up | 38025 |
Number of Leads Active in Study | 81 |
US Acute Lead Observations
Conductor Fracture | 2 |
---|---|
Failure to Capture | 17 |
Failure to Sense | 3 |
Impedance Out of Range | 10 |
Lead Dislodgement | 24 |
Oversensing | 18 |
Unspecified Clinical Failure | 6 |
Product Characteristics
Lead Type | Defibrillation |
---|---|
Type and/or Fixation | Tines |
Lead Function | Pacing/Sensing and Defibrillation |
Lead Placement | Transvenous |
Lead Tip Location | Right Ventricle |
Pace/Sense Polarity | True Bipolar/Two Coils |
Serial Number Prefix | TDC |
---|---|
Lead Body Diameter | 8.2 FR |
Insulation | Silicone with Polyurethane Overlay |
Pace/Sense Connector Style | IS-1 BI |
High Voltage Connector | DF-1 |
Steroid | Yes |
Data as of December 1, 2023