4968
CapSure Epi
ALL
Device Survival Probability
4968_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | 14 yr | |||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | 99.4 | 97.5 | 96.0 | 94.3 | 93.1 | 91.0 | 89.4 | 89.1 | 84.8 | 83.6 | 80.3 | 77.0 | 73.7 | 70.3 | ||||||||
# | 821 | 736 | 657 | 566 | 502 | 424 | 366 | 302 | 222 | 180 | 138 | 103 | 71 | 54 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 1050 |
---|---|
Cumulative Months of Follow-Up | 68092 |
Number of Leads Active in Study | 176 |
US Acute Lead Observations
Cardiac Perforation | 1 |
---|---|
Conductor Fracture | 4 |
Extra Cardiac Stimulation | 7 |
Failure to Capture | 88 |
Failure to Sense | 11 |
Impedance Out of Range | 21 |
Insulation Breach | 1 |
Lead Dislodgement | 8 |
Oversensing | 31 |
Product Characteristics
Lead Type | Epi/Myocardial |
---|---|
Type and/or Fixation | Suture |
Lead Function | Pacing/Sensing |
Lead Placement | Myocardial |
Lead Tip Location | Atrium or Right Ventricle |
Pace/Sense Polarity | Bipolar |
Serial Number Prefix | LEN |
---|---|
Lead Body Diameter | 7.8 FR |
Insulation | Silicone |
Pace/Sense Connector Style | IS-1 BI |
High Voltage Connector | n/a |
Steroid | Yes |
Data as of December 1, 2023