4194
Attain OTW
ALL
![Information](/productperformance-static/images/infow.png)
Device Survival Probability
4194_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | 14 yr | at 174.0 mo | ||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
98.6 | 97.4 | 96.7 | 96.1 | 95.6 | 94.3 | 94.1 | 93.5 | 93.5 | 93.3 | 92.9 | 92.9 | 92.9 | 91.1 | 91.1 | ||||||
# | 1238 | 1046 | 898 | 770 | 698 | 616 | 505 | 428 | 358 | 296 | 231 | 161 | 115 | 78 | 60 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 1655 |
---|---|
Cumulative Months of Follow-Up | 98854 |
Number of Leads Active in Study | 186 |
![Information](/productperformance-static/images/infow.png)
US Acute Lead Observations
Cardiac Perforation | 2 |
---|---|
Conductor Fracture | 2 |
Extra Cardiac Stimulation | 49 |
Failure to Capture | 42 |
Impedance Out of Range | 9 |
Lead Dislodgement | 153 |
Oversensing | 2 |
Unspecified Clinical Failure | 4 |
![Information](/productperformance-static/images/infow.png)
Product Characteristics
Lead Type | Left-Heart |
---|---|
Type and/or Fixation | Double Curve |
Lead Function | Pacing/Sensing |
Lead Placement | Transvenous |
Lead Tip Location | Left Ventricular Cardiac Vein |
Pace/Sense Polarity | Bipolar |
Serial Number Prefix | LFG |
---|---|
Lead Body Diameter | 6.0 FR |
Insulation | Polyurethane (55D) outer; silicone inner |
Pace/Sense Connector Style | IS-1 BI |
High Voltage Connector | n/a |
Steroid | Yes |
Data as of December 1, 2023