4194 Attain OTW
Left-Heart Pacing
4194 Attain OTW
ALL
Information

Device Survival Probability

4194_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr at 174.0 mo            
% Survival 98.6 97.4 96.7 96.1 95.6 94.3 94.1 93.5 93.5 93.3 92.9 92.9 92.9 91.1 91.1            
#   1238 1046 898 770 698 616 505 428 358 296 231 161 115 78 60            

Product Surveillance Registry Results

Number of Leads Enrolled in Study 1655
Cumulative Months of Follow-Up 98854
Number of Leads Active in Study 186
Information 68

Qualifying Complications

Conductor Fracture 2
Extra Cardiac Stimulation 11
Failure to Capture 22
Impedance Out of Range 0
Insulation (ESC) 1
Insulation (not further defined) 2
Lead Dislodgement 30
Other 0
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 2
Conductor Fracture 2
Extra Cardiac Stimulation 49
Failure to Capture 42
Impedance Out of Range 9
Lead Dislodgement 153
Oversensing 2
Unspecified Clinical Failure 4
Information

Distribution Data

US Market Release2004-08-24
CE Approval Date2003-07-14
Registered USA Implants114259
Estimated Active USA Implants28808
US Market Release2004-08-24
CE Approval Date2003-07-14
Estimated WW Distribution244145
Information

USA Returned Product Analysis

Conductor Fracture 48
Insulation Breach 165
Crimp/Weld/Bond 0
Other 2
Information

WW Returned Product Analysis

Conductor Fracture 54
Insulation Breach 171
Crimp/Weld/Bond 0
Other 4
Information

Product Characteristics

Lead Type Left-Heart
Type and/or Fixation Double Curve
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Left Ventricular Cardiac Vein
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LFG
Lead Body Diameter 6.0 FR
Insulation Polyurethane (55D) outer; silicone inner
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of December 1, 2023

www.medtronic.com/productperformance 951051