EN1SR01
Ensura SR MRI

Device Survival Probability
A3SR01_ENSR01_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | at 71.0 mo | ||||||
---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | ||||
% | ![]() |
99.9 | 99.9 | 99.8 | 99.7 | 99.7 | 99.7 | ||||
# | 23021 | 20141 | 17465 | 13236 | 5713 | 254 |
- Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
- Excluding Normal Battery Depletion – This is the malfunction free survival curve.

Product Characteristics
Generator Type | IPG |
---|---|
NBD Code | VVIR |
Max. Delivered Energy | N/A |
Connector Style | IS-1 UNI/BI |
---|---|
Serial Number Prefix | PTZ |
Mass | 22 g |
Volume | 12.7 cc |
Waveform | N/A |
X-Ray ID | PVX |
IS-1 UNI/BI

Estimated Longevity
Amplitude Setting | 500 Lead Ω | 1000 Lead Ω |
---|---|---|
Low 2.5 V (A, RV) | 11.9 | 12.9 |
Nominal 3.5 V (A, RV) | 10.0 | 11.6 |
High 5.0 V (A, RV) | 7.8 | 9.83 |
Longevity estimates based on the following device usage. Pace/Sense Mode VVI ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; EGM Prestorage ON ; Upper Sensing Rate 130 ppm; EGM Prestorage ON Time 1 ; Hide this content

Distribution Data
US Market Release | |
---|---|
CE Approval Date | 2014-04-24 |
Registered USA Implants | 1 |
Estimated Active USA Implants | 1 |
Normal Battery Depletions |
US Market Release | |
---|---|
CE Approval Date | 2014-04-24 |
Estimated WW Distribution | 64074 |
Normal Battery Depletions | 1 |

EOS Indication
From the point that the RRT is set, the pacemaker will operate for approximately six months for typical pacemaker configurations during the normal operating life.