4798 Attain Stability Quad
Left-Heart Pacing
4798 Attain Stability Quad
ALL
Information

Device Survival Probability

4798_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr at 54.0 mo          
% Survival 99.9 99.7 99.7 99.7 99.7          
#   597 314 142 74 62          

Product Surveillance Registry Results

Number of Leads Enrolled in Study 1190
Cumulative Months of Follow-Up 18836
Number of Leads Active in Study 872
Information 3

Qualifying Complications

Conductor Fracture 1
Extra Cardiac Stimulation 1
Failure to Capture 0
Impedance Out of Range 0
Lead Dislodgement 1
Other 0
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 7
Conductor Fracture 2
Extra Cardiac Stimulation 78
Failure to Capture 80
Impedance Out of Range 30
Lead Dislodgement 92
Oversensing 1
Information

Distribution Data

US Market Release2019-06-03
CE Approval Date2017-04-24
Registered USA Implants43449
Estimated Active USA Implants40722
US Market Release2019-06-03
CE Approval Date2017-04-24
Estimated WW Distribution115108
Information

USA Returned Product Analysis

Conductor Fracture 1
Insulation Breach 0
Crimp/Weld/Bond 0
Other 14
Information

WW Returned Product Analysis

Conductor Fracture 1
Insulation Breach 0
Crimp/Weld/Bond 0
Other 19
Information

Product Characteristics

Lead Type Left-Heart
Type and/or Fixation Non-electrically Active Side Fixation
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Left Ventricular Cardiac Vein
Pace/Sense Polarity Quadripolar
More
Serial Number Prefix QFX
Lead Body Diameter 4.4 FR
Insulation Polyurethane (outer) SI-polyimide (SI-PI)(inner)
Pace/Sense Connector Style IS4-LLLL
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of December 1, 2023

www.medtronic.com/productperformance 951051