4092
CapSure SP Novus
ALL

Device Survival Probability
4092_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | |||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
98.8 | 98.7 | 98.5 | 98.1 | 97.8 | 97.4 | 97.4 | 97.4 | 97.4 | 97.4 | 97.4 | 97.4 | |||||||||
# | 921 | 822 | 734 | 629 | 515 | 402 | 331 | 277 | 237 | 153 | 89 | 52 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 1200 |
---|---|
Cumulative Months of Follow-Up | 69410 |
Number of Leads Active in Study | 31 |

US Acute Lead Observations
Cardiac Perforation | 4 |
---|---|
Conductor Fracture | 4 |
Extra Cardiac Stimulation | 1 |
Failure to Capture | 35 |
Impedance Out of Range | 2 |
Insulation Breach | 1 |
Lead Dislodgement | 35 |
Oversensing | 1 |
Unspecified Clinical Failure | 1 |

Product Characteristics
Lead Type | Pacing |
---|---|
Type and/or Fixation | Tines |
Lead Function | Pacing/Sensing |
Lead Placement | Transvenous |
Lead Tip Location | Right Ventricle |
Pace/Sense Polarity | Bipolar |
Serial Number Prefix | LEP |
---|---|
Lead Body Diameter | 5.3 FR |
Insulation | Polyurethane (outer) Silicone (inner) |
Pace/Sense Connector Style | IS-1 BI |
High Voltage Connector | n/a |
Steroid | Yes |
Data as of October 29, 2020