4196 Attain Ability
Left-Heart Pacing
4196 Attain Ability
ALL
Information

Device Survival Probability

4196_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr
% Survival 98.0 97.3 96.6 96.0 95.5 94.8 94.5 94.2 93.2 92.7
#   1872 1483 1174 935 759 594 448 334 226 98

Product Surveillance Registry Results

Number of Leads Enrolled in Study 2303
Cumulative Months of Follow-Up 107956
Number of Leads Active in Study 376
Information 87

Qualifying Complications

Conductor Fracture 3
Extra Cardiac Stimulation 14
Failure to Capture 40
Impedance Out of Range 2
Insulation (not further defined) 1
Lead Dislodgement 23
Other 4
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 3
Conductor Fracture 2
Extra Cardiac Stimulation 95
Failure to Capture 64
Failure to Sense 1
Impedance Out of Range 10
Insulation Breach 1
Lead Dislodgement 224
Oversensing 1
Unspecified Clinical Failure 2
Information

Distribution Data

US Market Release2009-05-15
CE Approval Date2007-07-24
Registered USA Implants69830
Estimated Active USA Implants29438
US Market Release2009-05-15
CE Approval Date2007-07-24
Estimated WW Distribution211343
Information

USA Returned Product Analysis

Conductor Fracture 26
Insulation Breach 2
Crimp/Weld/Bond 0
Other 9
Information

WW Returned Product Analysis

Conductor Fracture 38
Insulation Breach 3
Crimp/Weld/Bond 0
Other 12
Information

Product Characteristics

Lead Type Left-Heart
Type and/or Fixation Double Curve
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Left Ventricular Cardiac Vein
Pace/Sense Polarity Bipolar
More
Serial Number Prefix PVI
Lead Body Diameter 4.0 FR
Insulation Polyurethane (outer); SI-polyimide (inner)
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of October 11, 2021

www.medtronic.com/productperformance 826050