6947M
Sprint Quattro Secure
ALL

Device Survival Probability
6947M_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | at 102.0 mo | |||
---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
99.7 | 99.5 | 99.4 | 99.4 | 99.0 | 99.0 | 98.1 | 97.9 | 97.9 | |
# | 1766 | 1491 | 1317 | 1115 | 940 | 752 | 603 | 308 | 140 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 2215 |
---|---|
Cumulative Months of Follow-Up | 111865 |
Number of Leads Active in Study | 781 |

US Acute Lead Observations
Cardiac Perforation | 39 |
---|---|
Conductor Fracture | 14 |
Extra Cardiac Stimulation | 12 |
Failure to Capture | 106 |
Failure to Sense | 40 |
Impedance Out of Range | 31 |
Lead Dislodgement | 224 |
Oversensing | 79 |

Product Characteristics
Lead Type | Defibrillation |
---|---|
Type and/or Fixation | Active Screw In |
Lead Function | Pacing/Sensing and Defibrillation |
Lead Placement | Transvenous |
Lead Tip Location | Right Ventricle |
Pace/Sense Polarity | True Bipolar/Two Coils |
Serial Number Prefix | TDK |
---|---|
Lead Body Diameter | 8.6 FR |
Insulation | Silicone with Polyurethane Overlay |
Pace/Sense Connector Style | DF4-LLHH |
High Voltage Connector | DF4-LLHH |
Steroid | Yes |
Data as of December 16, 2021