4296
Attain Ability Plus
ALL

Device Survival Probability
4296_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | at 102.0 mo | |||
---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
98.7 | 97.9 | 97.7 | 97.2 | 96.7 | 96.5 | 96.5 | 96.5 | 96.5 | |
# | 1155 | 930 | 760 | 634 | 531 | 447 | 331 | 151 | 89 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 1462 |
---|---|
Cumulative Months of Follow-Up | 67772 |
Number of Leads Active in Study | 361 |

US Acute Lead Observations
Cardiac Perforation | 2 |
---|---|
Conductor Fracture | 1 |
Extra Cardiac Stimulation | 61 |
Failure to Capture | 32 |
Impedance Out of Range | 11 |
Insulation Breach | 4 |
Lead Dislodgement | 118 |

Product Characteristics
Lead Type | Left-Heart |
---|---|
Type and/or Fixation | Double Curve |
Lead Function | Pacing/Sensing |
Lead Placement | Transvenous |
Lead Tip Location | Left Ventricular Cardiac Vein |
Pace/Sense Polarity | Dual Electrodes |
Serial Number Prefix | RRA |
---|---|
Lead Body Diameter | 5.3 FR |
Insulation | Polyurethane (outer) Silicone (inner) |
Pace/Sense Connector Style | IS-1 BI |
High Voltage Connector | n/a |
Steroid | Yes |
Data as of December 16, 2021