4074
CapSure Sense
VENTRICULAR PLACEMENT
Device Survival Probability
4074_VEN_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | 14 yr | 15 yr | 16 yr | 17 yr | at 210.0 mo | |||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | 99.4 | 99.3 | 99.2 | 99.2 | 99.2 | 99.2 | 99.2 | 99.2 | 99.2 | 98.8 | 98.8 | 97.6 | 97.6 | 96.8 | 96.8 | 96.8 | 96.8 | 96.8 | ||||
# | 1030 | 879 | 738 | 635 | 493 | 399 | 339 | 298 | 258 | 219 | 182 | 154 | 126 | 118 | 111 | 90 | 62 | 51 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 1192 |
---|---|
Cumulative Months of Follow-Up | 81246 |
Number of Leads Active in Study | 142 |
ATRIAL PLACEMENT
Device Survival Probability
4074_ATR_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | 14 yr | 15 yr | 16 yr | 17 yr | at 210.0 mo | |||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | ||||
# | 214 | 205 | 198 | 183 | 167 | 158 | 148 | 136 | 126 | 117 | 110 | 107 | 99 | 95 | 88 | 77 | 59 | 52 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 227 |
---|---|
Cumulative Months of Follow-Up | 29664 |
Number of Leads Active in Study | 56 |
US Acute Lead Observations
Cardiac Perforation | 36 |
---|---|
Conductor Fracture | 2 |
Extra Cardiac Stimulation | 4 |
Failure to Capture | 196 |
Failure to Sense | 19 |
Impedance Out of Range | 14 |
Lead Dislodgement | 213 |
Oversensing | 9 |
Product Characteristics
Lead Type | Pacing |
---|---|
Type and/or Fixation | Tines |
Lead Function | Pacing/Sensing |
Lead Placement | Transvenous |
Lead Tip Location | Right Ventricle |
Pace/Sense Polarity | Bipolar |
Serial Number Prefix | BBD |
---|---|
Lead Body Diameter | 5.3 FR |
Insulation | Polyurethane (outer) Silicone (inner) |
Pace/Sense Connector Style | IS-1 BI |
High Voltage Connector | n/a |
Steroid | Yes |
Data as of November 20, 2024