4074
CapSure Sense
VENTRICULAR PLACEMENT

Device Survival Probability
4074_VEN_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | 14 yr | at 174.0 mo | ||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
99.4 | 99.3 | 99.2 | 99.2 | 99.2 | 99.2 | 99.2 | 99.2 | 99.2 | 98.7 | 98.7 | 97.9 | 97.9 | 96.8 | 96.8 | ||||||
# | 1010 | 838 | 697 | 592 | 461 | 371 | 296 | 243 | 198 | 162 | 150 | 138 | 102 | 64 | 52 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 1177 |
---|---|
Cumulative Months of Follow-Up | 70949 |
Number of Leads Active in Study | 288 |
ATRIAL PLACEMENT

Device Survival Probability
4074_ATR_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | 14 yr | at 174.0 mo | ||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | ||||||
# | 214 | 205 | 198 | 183 | 167 | 158 | 148 | 136 | 127 | 117 | 106 | 103 | 86 | 59 | 51 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 227 |
---|---|
Cumulative Months of Follow-Up | 26032 |
Number of Leads Active in Study | 94 |

US Acute Lead Observations
Cardiac Perforation | 27 |
---|---|
Conductor Fracture | 2 |
Extra Cardiac Stimulation | 3 |
Failure to Capture | 126 |
Failure to Sense | 8 |
Impedance Out of Range | 4 |
Lead Dislodgement | 164 |
Oversensing | 7 |

Product Characteristics
Lead Type | Pacing |
---|---|
Type and/or Fixation | Tines |
Lead Function | Pacing/Sensing |
Lead Placement | Transvenous |
Lead Tip Location | Right Ventricle |
Pace/Sense Polarity | Bipolar |
Serial Number Prefix | BBD |
---|---|
Lead Body Diameter | 5.3 FR |
Insulation | Polyurethane (outer) Silicone (inner) |
Pace/Sense Connector Style | IS-1 BI |
High Voltage Connector | n/a |
Steroid | Yes |
Data as of October 29, 2020