Unipolar Longevity Estimation Software Error
Azure/Astra DR and SR Pacemakers and Percepta/Serena/Solara CRT-Ps
Original Date of Communication:
April 2021
View specific models that this applies to
STATUS UPDATE - NOVEMBER 2023
Through 26 October 2023, Medtronic has received 34 complaints from clinicians related to this issue. No permanent patient harms have been reported due to this issue.
ORIGINAL COMMUNICATION - APRIL 2021
This notice is to inform you of the availability of software updates to address a potential inaccurate longevity estimate that may occur with the Azure™ and Astra™ family of pacemakers (IPGs) and the Percepta™, Serena™, Solara™ family of cardiac resynchronization therapy pacemakers (CRT-Ps). A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted. For devices programmed to bipolar pacing in both the RA and RV chambers, the longevity estimates are not affected by this issue.
Through 05 March 2021, Medtronic has received 13 complaints from clinicians related to this issue. No permanent patient harms have been reported due to this issue. If the software update is not applied to the programmer, confusion regarding device longevity could lead to a missed RRT alert and a potential delayed intervention. Battery performance is not affected by this programmer display error. RRT will alert appropriately, and if patients are followed per standard clinical practice, the risk to patients is minimal.
The longevity estimation error associated with unipolar pacing configurations occurs only in the first half of the device life (prior to 50% depletion of the battery). During this phase, the estimator algorithm utilizes lead impedance as an input. When a unipolar pacing configuration is programmed, the algorithm incorrectly uses the bipolar lead impedance as input (rather than the appropriate unipolar pacing lead impedance). As a result, the algorithm will over-estimate the remaining longevity during this phase. At all times, RRT, Elective Replacement Indication, and End of Service will accurately display on programmers – even if the software update has not yet been installed.
Software updates are now available for Medtronic model 2090, Model 29901 Encore programmers, and SmartSync to correct this programmer display issue (see Table 1 below).
As of 26 March 2021, Medtronic CareLink network has been updated and will display correct longevity estimates for devices affected by this issue. Azure IPG and Percepta/Serena/Solara CRT-P patients remotely monitored via the MyCareLink Heart™ mobile app will automatically receive an updated longevity estimate on their mobile app with their next scheduled transmission or within 92 days of their last longevity update, whichever occurs first.
No change in patient management is necessary. Per labeling, the RRT notification can be used to identify when device replacement should be scheduled. There is no need to schedule patients to come in before their next regularly scheduled follow-up visit. The corrective fix for this error is implemented in programmers and the CareLink network. The patient’s implanted device does not require an update.
Based on your facility’s needs and accessibility, a Medtronic Representative or authorized personnel will assist with installing software on programmers in your account. Until all SmartSync and Model 2090 and Encore programmers are updated, a difference in longevity estimates may be displayed between programmers and the CareLink network due to this software error.