Device Survival Probability
SECURA_VR_SURV
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | at 122.0 mo | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
100.0 | 99.9 | 99.9 | 99.9 | 99.8 | 99.8 | 99.8 | 99.7 | 99.6 | 99.6 | 99.6 | ||||||||||
% | ![]() |
99.7 | 99.5 | 99.3 | 99.0 | 98.6 | 98.1 | 96.2 | 91.2 | 76.4 | 43.8 | 36.5 | ||||||||||
# | 17912 | 16706 | 15718 | 14644 | 13440 | 12196 | 10777 | 7948 | 4358 | 651 | 244 |
- Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
- Excluding Normal Battery Depletion – This is the malfunction free survival curve.

Product Characteristics
Generator Type | ICD |
---|---|
NBD Code | VVE-VVIR |
Max. Delivered Energy | 35 J |
Connector Style | Cx Style |
---|---|
Serial Number Prefix | PUX |
Mass | 68g |
Volume | 37cc |
Waveform | Biphasic |
X-Ray ID | PUG |


Estimated Longevity
Charging Frequency | 100% Pacing | 50% Pacing | 15% Pacing | 100% Sensing |
---|---|---|---|---|
Monthly | 4.3 | 4.7 | 4.9 | 5.0 |
Quarterly | 6.7 | 7.5 | 8.1 | 8.4 |
Semi-annual | 7.8 | 8.9 | 9.8 | 10.3 |
Longevity estimates based on the following device usage. Pacing Mode VVI; Ventricular Lower Pacing Rate 60ppm; Ventricular Pulse Width 0.4ms; Ventricular Pulse Amplitude 3V; Ventricle Impedance 510ohms; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Optivol ON; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Distribution Data
US Market Release | 2008-09-15 |
---|---|
CE Approval Date | |
Registered USA Implants | 20048 |
Estimated Active USA Implants | 6353 |
Normal Battery Depletions | 1766 |
US Market Release | 2008-09-15 |
---|---|
CE Approval Date | |
Estimated WW Distribution | 20470 |
Normal Battery Depletions | 1767 |

EOS Indication
The Prolonged Service Period (PSP) is the time between the RRT and EOS. The PSP is defined as 3 months assuming the following conditions: 100% VVI pacing at 60 min-1, 2.5 V RV pacing amplitude; 0.4 ms pulse width; 600 Ω pacing load; and 6 full-energy charges. The EOS may be indicated before the end of 3 months if the device exceeds these conditions.
Data as of October 29, 2020