SPDR01 Sphera DR MRI
Implantable Pulse Generator (IPG)
Device Survival Probability
Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Customer Communications For This Model
There are no communications for this model.
Product Characteristics
Generator Type
IPG
NBD Code
DDDR
Max. Delivered Energy
N/A
More
Connector Style
IS-1 UNI/BI
Serial Number Prefix
FNC
Mass
27.1 g
Volume
12.1 cc
Waveform
N/A
X-Ray ID
PHX
Hide this content
Estimated Longevity
Amplitude Setting
500 Lead Ω
1000 Lead Ω
Low 2.5 V (A, RV)
9.5
10.8
Nominal 3.5 V (A, RV)
7.4
9.5
High 5.0 V (A, RV)
5.2
7.3
USA
WW
Distribution Data
US Market Release 2017-08-03
CE Approval Date 2017-06-16
Registered USA Implants
Estimated Active USA Implants
Normal Battery Depletions
US Market Release 2017-08-03
CE Approval Date 2017-06-16
Estimated WW Distribution 65744
Normal Battery Depletions
Malfunctions (USA)
Therapy Function Not Compromised
Therapy Function Compromised
Malfunctions (WW)
Therapy Function Not Compromised
3
Electrical Component
3
Therapy Function Compromised
0
EOL Indicator
180 days from RRT
Recommended Replacement (RRT)
Battery voltage less than or equal to 2.5 V, or battery voltage less than or equal to 2.59 V and battery impedance at 3000 ohms
Data as of April 30, 2024