SESR01 Sensia SR
Implantable Pulse Generator (IPG)
SESR01 Sensia SR
Information

Device Survival Probability

SESR01_SES01_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr at 151.0 mo                
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0                
% Including Normal Battery Depletion 99.9 99.8 99.6 99.2 98.7 97.8 95.9 91.2 78.3 60.0 44.9 37.2 31.6                
#   85830 74457 64552 56011 48248 41084 34582 27602 18942 10422 4700 1561 151                
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

Product Characteristics

Generator Type IPG
NBD Code SSIR
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix PWR, NWR
Mass 21.5 g
Volume 9.7 cc
Waveform N/A
X-Ray ID PWL, PWB
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 7.4 | 8.9* 7.9 | 9.7*
Nominal 3.5 V (A, RV) 6.5 | 7.7* 7.5 | 9.1*
High 5.0 V (A, RV) 5.1 | 5.9* 6.3 | 7.4*
Footnote

*The second longevity numbers provided apply to the N-series Sensia SESR01 product. The N-series products have increased longevity because of improved electronic circuitry. Click on "Other Articles" for more information on the N-series longevity. Longevity estimates based on the following device usage. Pace/Sense Mode VVI ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; EGM Prestorage ON ; Upper Sensing Rate 130 ppm; EGM Prestorage ON Time 1 month; 100% pacing and pulse width 0.4 ms.Hide this content

Information

Distribution Data

US Market Release2006-07-17
CE Approval Date2005-09-20
Registered USA Implants117372
Estimated Active USA Implants22056
Normal Battery Depletions8574
US Market Release2006-07-17
CE Approval Date2005-09-20
Estimated WW Distribution550684
Normal Battery Depletions9710
Information

Malfunctions (USA)

Therapy Function Not Compromised 13
Electrical Component 7
Possible Early Battery Depletion 4
Other 2
Therapy Function Compromised 4
Electrical Component 3
Electrical Interconnect 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 26
Electrical Component 16
Possible Early Battery Depletion 7
Other 3
Therapy Function Compromised 12
Electrical Component 11
Electrical Interconnect 1
Information

EOS Indication

Prolonged service period is 3 months after RRT/ERI assuming 100% VVI/65 ppm pacing at the listed parameter settings. Reprogramming pacing parameters may reduce the duration of the prolonged service period.

Information

Recommended Replacement (RRT)

Telemetry indication. Rate and Mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of April 30, 2024

www.medtronic.com/productperformance 1001052