REVDD01
Relia VDD
Device Survival Probability
ADVDD01_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | at 113.0 mo | ||
---|---|---|---|---|---|---|---|---|---|---|---|
% | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | |
% | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 97.0 | 85.9 | 66.4 | 56.2 | 54.1 | |
# | 704 | 645 | 584 | 526 | 468 | 407 | 316 | 185 | 123 | 105 |
- Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
- Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Product Characteristics
Generator Type | IPG |
---|---|
NBD Code | VDD |
Max. Delivered Energy | N/A |
Connector Style | IS-1 UNI/BI |
---|---|
Serial Number Prefix | NWX |
Mass | 23.6 g |
Volume | 11.1 cc |
Waveform | N/A |
X-Ray ID | PWB |
IS-1 UNI/BI
Estimated Longevity
Amplitude Setting | 500 Lead Ω | 1000 Lead Ω |
---|---|---|
Low 2.5 V (A, RV) | 7.9 | 8.5 |
Nominal 3.5 V (A, RV) | 7.0 | 8.1 |
High 5.0 V (A, RV) | 5.4 | 6.7 |
Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; EGM Prestorage ON ; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; EGM Prestorage ON Time 1 month; Atrial Fib Percent of Time 0 %Hide this content
Distribution Data
US Market Release | |
---|---|
CE Approval Date | 2008-05-07 |
Registered USA Implants | 1 |
Estimated Active USA Implants | 0 |
Normal Battery Depletions |
US Market Release | |
---|---|
CE Approval Date | 2008-05-07 |
Estimated WW Distribution | 39509 |
Normal Battery Depletions | 22 |
EOS Indication
The prolonged service period is 3 months after RRT/ERI and assumes 100% pacing at VVI/65 and the listed parameter values. Reprogramming of parameters may shorten the duration of the prolonged service period.
Data as of April 30, 2024