RES01
Relia S
Device Survival Probability
ADSR01_ADSR03_ADSR06_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | at 135.0 mo | |||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | ||||||||||
% | 99.9 | 99.7 | 99.5 | 99.0 | 98.3 | 97.1 | 94.6 | 88.7 | 72.2 | 51.2 | 29.0 | 21.5 | ||||||||||
# | 72028 | 62944 | 55093 | 48012 | 41196 | 34864 | 28921 | 22170 | 13793 | 5968 | 1154 | 325 |
- Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
- Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Product Characteristics
Generator Type | IPG |
---|---|
NBD Code | AAI/VVI |
Max. Delivered Energy | N/A |
Connector Style | IS-1 UNI/BI |
---|---|
Serial Number Prefix | NWV |
Mass | 21.5 g |
Volume | 9.7 cc |
Waveform | N/A |
X-Ray ID | PWB |
IS-1 UNI/BI
Estimated Longevity
Amplitude Setting | 500 Lead Ω | 1000 Lead Ω |
---|---|---|
Low 2.5 V (A, RV) | 8.9 | 9.7 |
Nominal 3.5 V (A, RV) | 7.7 | 9.1 |
High 5.0 V (A, RV) | 5.9 | 7.4 |
Longevity estimates based on the following device usage. Pace/Sense Mode VVI ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; EGM Prestorage ON ; Upper Sensing Rate 130 ppm; EGM Prestorage ON Time 1 monthHide this content
Distribution Data
US Market Release | |
---|---|
CE Approval Date | 2008-05-07 |
Registered USA Implants | 3 |
Estimated Active USA Implants | 1 |
Normal Battery Depletions |
US Market Release | |
---|---|
CE Approval Date | 2008-05-07 |
Estimated WW Distribution | 91658 |
Normal Battery Depletions | 155 |
EOS Indication
The prolonged service period is 3 months after RRT/ERI and assumes 100% pacing at VVI/65 and the listed parameter values. Reprogramming of parameters may shorten the duration of the prolonged service period.
Data as of April 30, 2024