DVMB1D4 Evera MRI XT
Implantable Cardioverter Defibrillator (ICD)
DVMB1D4 Evera MRI XT
Information

Device Survival Probability

EVERA_VR_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr at 124.0 mo                    
% Excluding Normal Battery Depletion 100.0 100.0 99.9 99.8 99.8 99.6 99.6 99.5 99.4 99.3 99.3                    
% Including Normal Battery Depletion 100.0 99.9 99.7 99.6 99.4 99.1 98.9 98.7 98.3 97.7 97.5                    
#   52431 48794 45402 42247 39173 36002 33070 26184 13541 3207 436                    
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type ICD
NBD Code VVE-VVIR
Max. Delivered Energy 36 J
More
Connector Style DF4 VR Style
Serial Number Prefix PKZ
Mass 77 g
Volume 33 cc
Waveform Biphasic
X-Ray ID PFZ
Hide this content
DF4 VR Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 4.9 5.3 5.7 5.8
Quarterly 7.8 8.8 9.7 10.1
Semi-annual 8.5 9.7 10.7 11.3
Footnote

Longevity estimates based on the following device usage. Pacing Mode VVI; Ventricular Lower Pacing Rate 60ppm; Ventricular Pulse Width 0.4ms; Ventricular Pulse Amplitude 3V; Ventricle Impedance 510ohms; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Optivol ON; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formationHide this content

Information

Distribution Data

US Market Release2015-09-11
CE Approval Date
Registered USA Implants20550
Estimated Active USA Implants13817
Normal Battery Depletions18
US Market Release2015-09-11
CE Approval Date
Estimated WW Distribution24156
Normal Battery Depletions18
Information

Malfunctions (USA)

Therapy Function Not Compromised 16
Battery 12
Electrical Component 3
Other 1
Therapy Function Compromised 18
Battery 14
Device-Related Current Pathway 4
Information

Malfunctions (WW)

Therapy Function Not Compromised 16
Battery 12
Electrical Component 3
Other 1
Therapy Function Compromised 18
Battery 14
Device-Related Current Pathway 4
Information

EOS Indication

From the point that the RRT is set, the pacemaker will operate for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Recommended Replacement (RRT)

Battery voltage less than 2.73V on 3 consecutive daily automatic measurments.

Information

Charge Time

BLACKWELL_VR_CHRG
Loading...

Data as of April 30, 2024

www.medtronic.com/productperformance 1001052