DDPA2D1 Cobalt XT
Implantable Cardioverter Defibrillator (ICD)
DDPA2D1 Cobalt XT
Information

Device Survival Probability

COBALT_CROME_DR_SURV
Loading...










    1 yr 2 yr 3 yr at 45.0 mo            
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0            
% Including Normal Battery Depletion 100.0 99.9 99.9 99.9            
#   24487 13947 4335 178            
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type ICD
NBD Code DDE-DDDR
Max. Delivered Energy 40 J
More
Serial Number Prefix RSL
Mass 79 g
Volume 33.1 cc
Waveform Biphasic
X-Ray ID PLS
Hide this content
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 4.9 5.2 5.4 5.5
Quarterly 7.9 8.6 9.1 9.4
Semi-annual 9.4 10.3 11.1 11.4
Footnote

Longevity estimates based on the following device usage. Pacing Mode DDD; Atrial Lower Pacing Rate 60ppm; Ventricular Lower Pacing Rate 60ppm; Atrial Pulse Width 0.4ms; Ventricular Pulse Width 0.4ms; Atrial Pulse Amplitude 1.5V; Ventricular Pulse Amplitude 2V; Atrial Impedance 500ohms; Ventricle Impedance 500ohms; Atrial Sensing Rate 70bpm; Ventricle Sensing Rate 70bpm;Atrial Fib percentage of time 0%; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formationHide this content

Information

Distribution Data

US Market Release2020-04-23
CE Approval Date2019-12-18
Registered USA Implants5741
Estimated Active USA Implants5495
Normal Battery Depletions
US Market Release2020-04-23
CE Approval Date2019-12-18
Estimated WW Distribution24918
Normal Battery Depletions
Information

Malfunctions (USA)

Therapy Function Not Compromised 1
Electrical Component 1
Therapy Function Compromised 0
Information

Malfunctions (WW)

Therapy Function Not Compromised 2
Electrical Component 2
Therapy Function Compromised 0
Information

EOS Indicator

From the point that the RRT is set, the pacemaker will operate for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.8 V on 3 consecutive daily automatic measurements.

Data as of April 30, 2024

www.medtronic.com/productperformance 1001052