D354VRM Protecta XT VR
Implantable Cardioverter Defibrillator (ICD)
D354VRM Protecta XT VR
Information

Device Survival Probability

D3xxVRx_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr at 139.0 mo                  
% Excluding Normal Battery Depletion 100.0 100.0 100.0 99.9 99.9 99.9 99.8 99.8 99.7 99.7 99.7 99.7                  
% Including Normal Battery Depletion 99.9 99.9 99.7 99.5 99.2 98.8 97.4 92.9 79.7 56.8 48.1 45.8                  
#   25804 23984 22243 20574 19021 17447 15688 13081 8453 4069 1831 170                  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

Product Characteristics

Generator Type ICD
NBD Code VVE-VVIR
Max. Delivered Energy 35 J
More
Connector Style DF4 VR Style
Serial Number Prefix PSH
Mass 73g
Volume 38cc
Waveform Biphasic
X-Ray ID PSI
Hide this content
DF4 VR Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 4.3 4.7 4.9 5.0
Quarterly 6.7 7.5 8.1 8.4
Semi-annual 7.8 8.9 9.8 10.3
Footnote

Longevity estimates based on the following device usage. Pacing Mode VVI; Ventricular Lower Pacing Rate 60ppm; Ventricular Pulse Width 0.4ms; Ventricular Pulse Amplitude 3V; Ventricle Impedance 510ohms; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Optivol ON; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

Distribution Data

US Market Release
CE Approval Date2010-12-17
Registered USA Implants1
Estimated Active USA Implants0
Normal Battery Depletions
US Market Release
CE Approval Date2010-12-17
Estimated WW Distribution14934
Normal Battery Depletions6
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised 24
Battery 14
Electrical Component 7
Electrical Interconnect 1
Other 2
Therapy Function Compromised 10
Battery 10
Information

EOL Indicator

From the point that the RRT is set, the pacemaker will operate for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.63V on 3 consecutive daily automatic measurments.

Information

Charge Time

D354VRx_CHRG
Loading...

Data as of April 30, 2024

www.medtronic.com/productperformance 1001052