C2TR01 Syncra CRT-P
Cardiac Resynchronization Therapy - Pacing (CRT-Pacing)
C2TR01 Syncra CRT-P
Information

Device Survival Probability

C2TR01_C3TR01_C4TR01_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr at 124.0 mo                    
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 99.9 99.9 99.9 99.9 99.9 99.9 99.9                    
% Including Normal Battery Depletion 99.9 99.8 99.5 98.0 95.5 91.2 84.5 74.1 60.2 43.1 31.0                    
#   26186 23390 20951 18301 15665 13084 10292 6908 3651 958 218                    
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

Product Characteristics

Generator Type CRT-Pacing
NBD Code OOE-DDDR
Max. Delivered Energy N/A
More
Connector Style CRT-P Style
Serial Number Prefix PZX
Mass 26 g
Volume 15 cc
Waveform N/A
X-Ray ID PVX
Hide this content
CRT-P Style
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 8.7 10.7
Nominal 3.5 V (A, RV) 6.0 8.2
High 5.0 V (A, RV) 3.3 5.1
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Left Ventricle Pulse Width 0.4 ms; Left Ventricle Pulse Amplitude same as A, RV; Left Ventricle Percent Paced 100 %; Left Ventricle Lead Impedance same as A, RV; EGM Prestorage ON ; EGM Prestorage ON Time 1 ; Atrial Fib Percent of Time 0 %Hide this content

Information

Distribution Data

US Market Release2011-03-22
CE Approval Date2010-05-11
Registered USA Implants10237
Estimated Active USA Implants2204
Normal Battery Depletions973
US Market Release2011-03-22
CE Approval Date2010-05-11
Estimated WW Distribution57229
Normal Battery Depletions983
Information

Malfunctions (USA)

Therapy Function Not Compromised 6
Possible Early Battery Depletion 5
Other 1
Therapy Function Compromised 1
Possible Early Battery Depletion 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 11
Electrical Component 3
Possible Early Battery Depletion 6
Software/Firmware 1
Other 1
Therapy Function Compromised 2
Possible Early Battery Depletion 1
Other 1
Information

EOS Indication

From the point that the RRT is set, the pacemaker will operate for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.77V on 3 consecutive daily automatic measurments.

Data as of April 30, 2024

www.medtronic.com/productperformance 1001052