ADSR03 Adapta SR
Implantable Pulse Generator (IPG)
ADSR03 Adapta SR
Information

Device Survival Probability

ADSR01_ADSR03_ADSR06_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr at 135.0 mo                  
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0                  
% Including Normal Battery Depletion 99.9 99.7 99.5 99.0 98.3 97.1 94.6 88.7 72.2 51.2 29.0 21.5                  
#   72028 62944 55093 48012 41196 34864 28921 22170 13793 5968 1154 325                  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

Product Characteristics

Generator Type IPG
NBD Code SSIR
Max. Delivered Energy N/A
More
Connector Style 3.2mm Long Pin
Serial Number Prefix NWP, PWP
Mass 22.5 g
Volume 10.5 cc
Waveform N/A
X-Ray ID PWB
Hide this content
3.2mm Long Pin
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 7.4 | 8.9* 7.9 | 9.7*
Nominal 3.5 V (A, RV) 6.5 | 7.7* 7.5 | 9.1*
High 5.0 V (A, RV) 5.1 | 5.9* 6.3 | 7.4*
Footnote

*The second longevity numbers provided apply to the N-series Adapta ADSR03 product. The N-series products have increased longevity because of improved electronic circuitry. Click on "Other Articles" for more information on the N-series longevity. Longevity estimates based on the following device usage. Pace/Sense Mode VVI ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; EGM Prestorage ON ; Upper Sensing Rate 130 ppm; EGM Prestorage ON Time 1 ; Hide this content

Information

Distribution Data

US Market Release2006-07-17
CE Approval Date2005-09-20
Registered USA Implants2119
Estimated Active USA Implants437
Normal Battery Depletions198
US Market Release2006-07-17
CE Approval Date2005-09-20
Estimated WW Distribution27854
Normal Battery Depletions851
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised 1
Electrical Component 1
Therapy Function Compromised 1
Electrical Component 1
Information

EOS Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Recommended Replacement (RRT)

Telemetry indication. Rate and mode change to 65 ppm and VVI respectively (VOO/65 with magnet).

Data as of April 30, 2024

www.medtronic.com/productperformance 1001052