A2DR01 Advisa DR MRI
Implantable Pulse Generator (IPG)
A2DR01 Advisa DR MRI
Information

Device Survival Probability

A2DR01_A3DR01_A5DR01_EN1DR01_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr at 116.0 mo
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0
% Including Normal Battery Depletion 100.0 99.9 99.9 99.8 99.7 99.4 98.9 95.9 87.6 55.5
#   308538 290678 273742 256730 237337 215434 151447 83731 27980 1468
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

Product Characteristics

Generator Type IPG
NBD Code OAE-DDDR
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix PVY
Mass 22 g
Volume 12.7 cc
Waveform N/S
X-Ray ID PVX
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 8.9 10.2
Nominal 3.5 V (A, RV) 6.9 8.6
High 5.0 V (A, RV) 4.3 6.2
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; EGM Prestorage ON ; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; EGM Prestorage ON Time 1 ; Atrial Fib Percent of Time 0 % Hide this content

Information

Distribution Data

US Market Release2013-01-15
CE Approval Date
Registered USA Implants344413
Estimated Active USA Implants217681
Normal Battery Depletions6817
US Market Release2013-01-15
CE Approval Date
Estimated WW Distribution351077
Normal Battery Depletions6821
Information

Malfunctions (USA)

Therapy Function Not Compromised 75
Battery 1
Electrical Component 38
Electrical Interconnect 4
Possible Early Battery Depletion 23
Software/Firmware 6
Other 3
Therapy Function Compromised 5
Electrical Component 5
Information

Malfunctions (WW)

Therapy Function Not Compromised 78
Battery 1
Electrical Component 41
Electrical Interconnect 4
Possible Early Battery Depletion 23
Software/Firmware 6
Other 3
Therapy Function Compromised 5
Electrical Component 5
Information

EOS Indication

From the point that the RRT is set, the pacemaker will operate for approximately six months for typical pacemaker configurations during the normal operating life.

Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.83V on 3 consecutive daily automatic measurments.

Data as of April 30, 2024

www.medtronic.com/productperformance 1001052