6947M Sprint Quattro Secure
Defibrillation
6947M Sprint Quattro Secure
ALL
Information

Device Survival Probability

6947M_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr                    
% Survival 99.7 99.5 99.4 99.4 99.0 98.9 98.2 97.9 97.4 97.1 97.1                    
#   1862 1554 1365 1163 1003 848 713 595 476 315 81                    

Product Surveillance Registry Results

Number of Leads Enrolled in Study 2410
Cumulative Months of Follow-Up 132529
Number of Leads Active in Study 671
Information 28

Qualifying Complications

Conductor Fracture 15
Failure to Capture 4
Failure to Sense 4
Impedance Out of Range 1
Lead Dislodgement 1
Oversensing 2
Other 1
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 40
Conductor Fracture 15
Extra Cardiac Stimulation 12
Failure to Capture 124
Failure to Sense 47
Impedance Out of Range 36
Insulation Breach 1
Lead Dislodgement 241
Oversensing 86
Information

Distribution Data

US Market Release2012-02-13
CE Approval Date2010-03-12
Registered USA Implants137519
Estimated Active USA Implants94732
US Market Release2012-02-13
CE Approval Date2010-03-12
Estimated WW Distribution283714
Information

USA Returned Product Analysis

Conductor Fracture 257
Insulation Breach 15
Crimp/Weld/Bond 1
Other 37
Information

WW Returned Product Analysis

Conductor Fracture 392
Insulation Breach 21
Crimp/Weld/Bond 1
Other 43
Information

Product Characteristics

Lead Type Defibrillation
Type and/or Fixation Active Screw In
Lead Function Pacing/Sensing and Defibrillation
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity True Bipolar/Two Coils
More
Serial Number Prefix TDK
Lead Body Diameter 8.6 FR
Insulation Silicone with Polyurethane Overlay
Pace/Sense Connector Style DF4-LLHH
High Voltage Connector DF4-LLHH
Steroid Yes
Hide this content

Data as of April 30, 2024

www.medtronic.com/productperformance 1001052