6947M
Sprint Quattro Secure
ALL
Device Survival Probability
6947M_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | 99.7 | 99.5 | 99.4 | 99.4 | 99.0 | 98.9 | 98.2 | 97.9 | 97.4 | 97.1 | 97.1 | |||||||||||
# | 1862 | 1554 | 1365 | 1163 | 1003 | 848 | 713 | 595 | 476 | 315 | 81 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 2410 |
---|---|
Cumulative Months of Follow-Up | 132529 |
Number of Leads Active in Study | 671 |
US Acute Lead Observations
Cardiac Perforation | 40 |
---|---|
Conductor Fracture | 15 |
Extra Cardiac Stimulation | 12 |
Failure to Capture | 124 |
Failure to Sense | 47 |
Impedance Out of Range | 36 |
Insulation Breach | 1 |
Lead Dislodgement | 241 |
Oversensing | 86 |
Product Characteristics
Lead Type | Defibrillation |
---|---|
Type and/or Fixation | Active Screw In |
Lead Function | Pacing/Sensing and Defibrillation |
Lead Placement | Transvenous |
Lead Tip Location | Right Ventricle |
Pace/Sense Polarity | True Bipolar/Two Coils |
Serial Number Prefix | TDK |
---|---|
Lead Body Diameter | 8.6 FR |
Insulation | Silicone with Polyurethane Overlay |
Pace/Sense Connector Style | DF4-LLHH |
High Voltage Connector | DF4-LLHH |
Steroid | Yes |
Data as of April 30, 2024