6947 Sprint Quattro Secure
Defibrillation
6947 Sprint Quattro Secure
ALL
Information

Device Survival Probability

6947_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr 15 yr 16 yr 17 yr 18 yr 19 yr at 234.0 mo  
% Survival 99.5 99.3 99.0 98.7 98.2 97.9 97.5 97.1 96.7 96.1 95.8 95.5 95.0 94.5 94.1 92.0 92.0 90.5 88.4 87.1  
#   3302 2902 2548 2259 2025 1787 1541 1380 1233 1073 885 679 514 332 204 146 120 100 78 58  

Product Surveillance Registry Results

Number of Leads Enrolled in Study 4583
Cumulative Months of Follow-Up 298935
Number of Leads Active in Study 570
Information 105

Qualifying Complications

Cardiac Perforation 1
Conductor Fracture 42
Failure to Capture 8
Failure to Sense 2
Impedance Out of Range 13
Insulation (not further defined) 6
Lead Dislodgement 5
Oversensing 21
Other 4
Unspecified Clinical Failure 3
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 29
Conductor Fracture 26
Extra Cardiac Stimulation 2
Failure to Capture 83
Failure to Sense 36
Impedance Out of Range 61
Insulation Breach 4
Lead Dislodgement 124
Oversensing 141
Unspecified Clinical Failure 20
Information

Distribution Data

US Market Release2001-11-12
CE Approval Date2001-10-04
Registered USA Implants375589
Estimated Active USA Implants125490
US Market Release2001-11-12
CE Approval Date2001-10-04
Estimated WW Distribution617061
Information

USA Returned Product Analysis

Conductor Fracture 1414
Insulation Breach 103
Crimp/Weld/Bond 4
Other 197
Information

WW Returned Product Analysis

Conductor Fracture 1749
Insulation Breach 134
Crimp/Weld/Bond 4
Other 239
Information

Product Characteristics

Lead Type Defibrillation
Type and/or Fixation Active Screw In
Lead Function Pacing/Sensing and Defibrillation
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity True Bipolar/Two Coils
More
Serial Number Prefix TDG
Lead Body Diameter 8.6 FR
Insulation Silicone with Polyurethane Overlay
Pace/Sense Connector Style IS-1 BI
High Voltage Connector DF-1
Steroid Yes
Hide this content

Data as of April 30, 2024

www.medtronic.com/productperformance 1001052