6935M Sprint Quattro Secure S
Defibrillation
6935M Sprint Quattro Secure S
ALL
Information

Device Survival Probability

6935M_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr
% Survival 99.7 99.5 99.2 99.0 98.6 98.2 97.7 97.3 96.8 96.8
#   7128 5483 4436 3648 2963 2343 1589 846 390 126

Product Surveillance Registry Results

Number of Leads Enrolled in Study 9745
Cumulative Months of Follow-Up 396299
Number of Leads Active in Study 4568
Information 111

Qualifying Complications

Cardiac Perforation 2
Conductor Fracture 50
Extra Cardiac Stimulation 1
Failure to Capture 14
Failure to Sense 1
Impedance Out of Range 10
Insulation (not further defined) 3
Lead Dislodgement 20
Oversensing 5
Other 4
Unspecified Clinical Failure 1
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 190
Conductor Fracture 22
Extra Cardiac Stimulation 32
Failure to Capture 453
Failure to Sense 153
Impedance Out of Range 130
Insulation Breach 3
Lead Dislodgement 663
Oversensing 350
Information

Distribution Data

US Market Release2012-08-02
CE Approval Date2012-07-12
Registered USA Implants390619
Estimated Active USA Implants320954
US Market Release2012-08-02
CE Approval Date2012-07-12
Estimated WW Distribution902508
Information

USA Returned Product Analysis

Conductor Fracture 785
Insulation Breach 35
Crimp/Weld/Bond 2
Other 102
Information

WW Returned Product Analysis

Conductor Fracture 1302
Insulation Breach 52
Crimp/Weld/Bond 3
Other 161
Information

Product Characteristics

Lead Type Defibrillation
Type and/or Fixation Active Screw In
Lead Function Pacing/Sensing and Defibrillation
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity True Bipolar/One Coil
More
Serial Number Prefix TDL
Lead Body Diameter 2.8 mm (8.4 Fr)
Insulation Silicone with Polyurethane Overlay
Pace/Sense Connector Style DF4-LLHO
High Voltage Connector DF4-LLHO
Steroid Yes
Hide this content

Data as of April 30, 2024