6935
Sprint Quattro Secure S
ALL
Device Survival Probability
6935_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | at 162.0 mo | |||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | 99.5 | 99.3 | 98.9 | 98.7 | 98.5 | 98.1 | 97.6 | 96.8 | 95.9 | 94.8 | 94.5 | 94.2 | 92.8 | 92.8 | ||||||||
# | 2443 | 1997 | 1639 | 1342 | 1139 | 980 | 817 | 695 | 587 | 470 | 316 | 178 | 90 | 60 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 3006 |
---|---|
Cumulative Months of Follow-Up | 168788 |
Number of Leads Active in Study | 671 |
US Acute Lead Observations
Cardiac Perforation | 30 |
---|---|
Conductor Fracture | 3 |
Extra Cardiac Stimulation | 2 |
Failure to Capture | 39 |
Failure to Sense | 15 |
Impedance Out of Range | 30 |
Insulation Breach | 1 |
Lead Dislodgement | 68 |
Oversensing | 68 |
Unspecified Clinical Failure | 5 |
Product Characteristics
Lead Type | Defibrillation |
---|---|
Type and/or Fixation | Active Screw In |
Lead Function | Pacing/Sensing and Defibrillation |
Lead Placement | Transvenous |
Lead Tip Location | Right Ventricle |
Pace/Sense Polarity | True Bipolar/One Coil |
Serial Number Prefix | TAU |
---|---|
Lead Body Diameter | 8,2 FR |
Insulation | Silicone with Polyurethane Overlay |
Pace/Sense Connector Style | IS-1 BI |
High Voltage Connector | DF-1 |
Steroid | Yes |
Data as of April 30, 2024