5092
CapSure SP Novus
ALL
Device Survival Probability
5092_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | at 162.0 mo | |||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | 99.5 | 99.3 | 99.2 | 98.9 | 98.9 | 98.6 | 98.6 | 98.6 | 98.6 | 98.6 | 97.8 | 97.8 | 97.8 | 97.8 | ||||||||
# | 814 | 652 | 517 | 421 | 335 | 265 | 218 | 173 | 149 | 133 | 109 | 81 | 57 | 52 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 1218 |
---|---|
Cumulative Months of Follow-Up | 54677 |
Number of Leads Active in Study | 15 |
US Acute Lead Observations
Cardiac Perforation | 7 |
---|---|
Conductor Fracture | 3 |
Extra Cardiac Stimulation | 3 |
Failure to Capture | 49 |
Failure to Sense | 7 |
Impedance Out of Range | 1 |
Insulation Breach | 3 |
Lead Dislodgement | 72 |
Oversensing | 1 |
Unspecified Clinical Failure | 8 |
Product Characteristics
Lead Type | Pacing |
---|---|
Type and/or Fixation | Tines |
Lead Function | Pacing/Sensing |
Lead Placement | Transvenous |
Lead Tip Location | Right Ventricle |
Pace/Sense Polarity | Bipolar |
Serial Number Prefix | LET |
---|---|
Lead Body Diameter | 6.0 FR |
Insulation | Silicone |
Pace/Sense Connector Style | IS-1 BI |
High Voltage Connector | n/a |
Steroid | Yes |
Data as of April 30, 2024