4968 CapSure Epi
Epi/Myocardial Pacing
4968 CapSure Epi
ALL
Information

Device Survival Probability

4968_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr 15 yr            
% Survival 99.4 97.5 96.0 94.3 93.1 91.1 89.3 89.1 85.0 83.9 80.6 77.9 75.1 72.3 71.3            
#   824 742 662 572 509 434 384 321 239 198 162 122 88 69 53            

Product Surveillance Registry Results

Number of Leads Enrolled in Study 1053
Cumulative Months of Follow-Up 70996
Number of Leads Active in Study 186
Information 106

Qualifying Complications

Conductor Fracture 33
Extra Cardiac Stimulation 2
Failure to Capture 29
Failure to Sense 3
Impedance Out of Range 5
Insulation (not further defined) 4
Lead Dislodgement 1
Oversensing 26
Other 3
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 1
Conductor Fracture 4
Extra Cardiac Stimulation 7
Failure to Capture 97
Failure to Sense 12
Impedance Out of Range 21
Insulation Breach 1
Lead Dislodgement 8
Oversensing 31
Information

Distribution Data

US Market Release1999-09-16
CE Approval Date1998-04-21
Registered USA Implants62709
Estimated Active USA Implants34288
US Market Release1999-09-16
CE Approval Date1998-04-21
Estimated WW Distribution225654
Information

USA Returned Product Analysis

Conductor Fracture 150
Insulation Breach 84
Crimp/Weld/Bond 0
Other 1
Information

WW Returned Product Analysis

Conductor Fracture 188
Insulation Breach 105
Crimp/Weld/Bond 0
Other 1
Information

Product Characteristics

Lead Type Epi/Myocardial
Type and/or Fixation Suture
Lead Function Pacing/Sensing
Lead Placement Myocardial
Lead Tip Location Atrium or Right Ventricle
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LEN
Lead Body Diameter 7.8 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of April 30, 2024