4598
Attain Performa S
ALL
Device Survival Probability
4598_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | at 90.0 mo | ||||
---|---|---|---|---|---|---|---|---|---|---|---|
% | 99.2 | 99.0 | 98.9 | 98.6 | 98.5 | 98.5 | 98.5 | 97.6 | |||
# | 1141 | 959 | 793 | 657 | 456 | 263 | 138 | 95 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 1375 |
---|---|
Cumulative Months of Follow-Up | 61197 |
Number of Leads Active in Study | 538 |
US Acute Lead Observations
Cardiac Perforation | 11 |
---|---|
Conductor Fracture | 2 |
Extra Cardiac Stimulation | 138 |
Failure to Capture | 100 |
Impedance Out of Range | 36 |
Lead Dislodgement | 86 |
Oversensing | 1 |
Product Characteristics
Lead Type | Left-Heart |
---|---|
Type and/or Fixation | S-shape |
Lead Function | Pacing/Sensing |
Lead Placement | Transvenous |
Lead Tip Location | Left Ventricular Cardiac Vein |
Pace/Sense Polarity | Quadripolar |
Serial Number Prefix | QUC |
---|---|
Lead Body Diameter | 5.3 FR |
Insulation | Polyurethane (outer) SI-polyimide (SI-PI)(inner) |
Pace/Sense Connector Style | IS4-LLLL |
High Voltage Connector | n/a |
Steroid | Yes |
Data as of April 30, 2024