4298
Attain Performa
ALL
Device Survival Probability
4298_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | at 102.0 mo | |||
---|---|---|---|---|---|---|---|---|---|---|---|
% | 99.2 | 98.7 | 98.6 | 98.6 | 98.4 | 98.2 | 98.2 | 98.2 | 98.2 | ||
# | 1882 | 1595 | 1363 | 1120 | 804 | 589 | 378 | 176 | 85 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 2262 |
---|---|
Cumulative Months of Follow-Up | 107609 |
Number of Leads Active in Study | 850 |
US Acute Lead Observations
Cardiac Perforation | 7 |
---|---|
Conductor Fracture | 1 |
Extra Cardiac Stimulation | 233 |
Failure to Capture | 168 |
Failure to Sense | 1 |
Impedance Out of Range | 44 |
Lead Dislodgement | 256 |
Product Characteristics
Lead Type | Left-Heart |
---|---|
Type and/or Fixation | Double Curve |
Lead Function | Pacing/Sensing |
Lead Placement | Transvenous |
Lead Tip Location | Left Ventricular Cardiac Vein |
Pace/Sense Polarity | Quadripolar |
Serial Number Prefix | QUA |
---|---|
Lead Body Diameter | 5.3 FR |
Insulation | Polyurethane (outer) Silicone (inner) |
Pace/Sense Connector Style | IS-4 |
High Voltage Connector | n/a |
Steroid | Yes |
Data as of April 30, 2024