4196
Attain Ability
ALL
Device Survival Probability
4196_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | 98.0 | 97.3 | 96.6 | 95.9 | 95.7 | 95.2 | 94.8 | 94.2 | 93.9 | 93.6 | 93.6 | 93.1 | 93.1 | |||||||||
# | 1887 | 1499 | 1192 | 964 | 792 | 636 | 495 | 410 | 339 | 262 | 194 | 134 | 51 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 2323 |
---|---|
Cumulative Months of Follow-Up | 119377 |
Number of Leads Active in Study | 219 |
US Acute Lead Observations
Cardiac Perforation | 3 |
---|---|
Conductor Fracture | 2 |
Extra Cardiac Stimulation | 98 |
Failure to Capture | 67 |
Failure to Sense | 1 |
Impedance Out of Range | 12 |
Insulation Breach | 1 |
Lead Dislodgement | 227 |
Oversensing | 1 |
Unspecified Clinical Failure | 2 |
Product Characteristics
Lead Type | Left-Heart |
---|---|
Type and/or Fixation | Double Curve |
Lead Function | Pacing/Sensing |
Lead Placement | Transvenous |
Lead Tip Location | Left Ventricular Cardiac Vein |
Pace/Sense Polarity | Bipolar |
Serial Number Prefix | PVI |
---|---|
Lead Body Diameter | 4.0 FR |
Insulation | Polyurethane (outer); SI-polyimide (inner) |
Pace/Sense Connector Style | IS-1 BI |
High Voltage Connector | n/a |
Steroid | Yes |
Data as of April 30, 2024